puxiliangyi@163.com

中文
Portable Shore Hardness Tester

News Center

current position: Home > News Center > Industry news

News CenterNews

contact usContact Us

PuXi Hardness Co.Ltd

Consumer hotline:

135-2128-6239;136-9117-2245

Mail:

puxiliangyi@163.com

Business Cooperation/Friendly Link:

vx a13041108151

WeChat Customer Service:

Yifan-1369

Company address:

101B, Block A, Jiande Commercial Building, No.2, Qijiahuozi A, Haidian District, Beijing


Shore hardness tester requirements for test specimens

2020-04-11 11:20:51
times

At present, the most commonly used Shore hardness tester in China is two pointer type Shore hardness tester


1 Shaw type A hardness tester


Shaw D type hardness tester


When using Shore hardness timing, A type A hardness tester reading below 10HA is inaccurate and the measurement results cannot be used. Type D hardness tester is recommended when the measured value of type A hardness tester exceeds 90HA. However, due to economic considerations, users generally only buy hardness tester and do not buy a fixed load frame, so novice operation is often not standard, resulting in a large deviation of the test results. How to use these two kinds of hardness tester correctly? In combination with the actual situation, the following methods are provided for the reference of test users


1. The surface of rubber and plastic samples should be smooth and flat, without mechanical damage and impurities and other defects.


2. Plastic samples and test temperature requirements;


(1) The plastic sample is square, with a side length of 50mm and a thickness of 6mm. Samples of 50×15mm are also allowed.


② Where possible, the samples shall be adjusted at laboratory standard temperature according to GB/T2941-1991 before testing. Comparison tests or series of tests must be carried out at the same temperature.


3. Rubber sample and test temperature requirements;


①, rubber sample thickness is not less than 6mm, width is not less than 15mm, length is not less than 35mm, sample thickness is less than 6mm, the same film can be overlapped determination, but not more than 3 layers. The film is parallel to the top and bottom.


② The room temperature was 23℃±5℃ during verification, and the durometer should be stored at this temperature for at least 1 hour before verification


AM型邵氏硬度计


Recently Viewed:

Related products

news.related_news

Beijing Spectrum Analyzer Technology Co., Ltd.

An enterprise specializing in R&D, production and sales of testing equipment and technical services

Contact us

135-2128-6329

Copyright © PuXi Hardness Co.Ltd All rights reserved record number:Beijing ICP No. 19033643 Mainly engaged in, Welcome to inquire!
Technical Support: 驰正科技
Main business area: 上海 北京 苏州 广州 深圳
Site map | RSS | XML